54

OUTPUT MIDDLE RANGE

To easily adjust a recorder offset, the analog output could be
set to middle range. In this case the output is set to 12 mA or
10 mA for 4-20 mA or 0-20 mA output.

To activate this option select “General Menu” - “Analog
Output1” or “Analog Output2” - “Output Middle Range”
function and press “CFM” key. The analyzer will display
“Recorder output is set to middle value”. Press “ESC” to exit
from this function.