21

Note: The electrode should be submerged

approximately 4 cm (1½") into the
solution with the temperature probe
located as close to it as possible.

• Select the second buffer value on the secondary display by

pressing the

Ï

°C or

Ð

CCCCC key.

• When the “READY” and “CON” symbols blink on the display,

the reading is stable and the calibration can be confirmed.

• Press the CFM key. If the reading is not close to the selected

buffer solution, “WRONG

2

BUF

” and “WRONG ” will blink

alternatively. If the reading is close to the selected buffer, the
slope and the offset are calibrated. The values will be stored in
memory and the meter will return to the operating mode.

Note: The meter will automatically skip the buffer that was used

for the first calibration to avoid erroneous calibration. At
least 1 full pH unit is required between the two buffers used
for the offset and slope calibration. As a result, once
calibrated at either pH 6.86 or 7.01, the microprocessor will
automatically ignore the other one for the second point
calibration. It will do likewise for pH 9.18 and 10.01

Note: During calibration, the secondary LCD displays the selected

buffer value. With HI 9025C, press the RANGE key to display
the buffer temperature during calibration.

PROCEDURE FOR HI 9110, HI 9210 & HI 9210N:

Manual Temperature Compensation
Note: The pH/ C electrode of HI 9210N incorporates a tempera-

ture sensor which performs automatic temperature

compensation. Consequently, the following 5 paragraphs
are irrelevant to

HI 9210N and you can just proceed with

pH calibration below.

READY

CON

CFM

WRONG

WRONG

2

BUF

pH

°C

SET

RANGE

°C

22

• Press pH/ C SET key for

HI 9110 or

the RANGE key for

HI 9210 to select the

manual temperature setting. The " C"
symbol will be displayed to indicate the
temperature setting mode.

• Remove the protective cap on the top of the meter to have

access to the calibration trimmers.

• Record the temperature of the buffer with a Checktemp C or

an accurate thermometer (e.g.20 C).

• Using the screwdriver, turn the temperature trimmer (# 5

pages 9-10) to display the recorded temperature value.

• Press the pH/ C SET key for

HI 9110

or the RANGE key for

HI 9210 to read

pH.

pH Calibration
All three meters provide for manual pH cali-

bration. As a result, you can perform
one or two point calibration at a buffer
of your choice. The instructions below
explain a typical calibration procedure at
pH 7 and then 4 or 10.

• Remove the protective cap from the elec-

trode, rinse it with some pH 7.01 solution, then dip in pH
7.01 buffer. Stir briefly and wait for one minute for the
reading to stabilize.

Note: The electrode should be submerged approximately 4 cm

°C

°C

pH

°C

SET

RANGE

°C

HI 7007

pH