Hanna Instruments HI 9023 User Manual

Page 9

Advertising
background image

17

CFM

Note: The electrode should be submerged

approximately 4 cm (1½") into the
solution. With the temperature probe
located as close to it as possible (HI

HI

HI

HI

HI

9023C

9023C

9023C

9023C

9023C).

• Press the CAL key. The temperature compensated pH value will

blink on the display. For example, if the temperature is 25 C the
reading will be pH 7.01. If the temperature is 20 C the reading

will be 7.03. The pH/temperature con-
version chart is on page 24.

Note: If "E4" flashes, the buffer solution is

wrong or the probe is defective (see page 25 for trouble-
shooting).

• Wait for the "pH" symbol to stop flashing and press the CFM

key. The symbol "E5" will flash to indicate that the electrode is
still in the pH 7.01 buffer.

• Rinse the pH electrode (and the tem-

perature probe) with distilled water first
and with the pH 4.01 (or 10.01) buffer
after.

• Dip the electrode (and the temperature

probe) into the pH 4.01 (or 10.01) buffer solution. Stir gently
and wait for the reading to stabilize.

• The "E5" symbol should disappear and the temperature com-

pensated value will blink on the display.
For example, if you are using pH 4.01 as
the buffer, the reading will be 4.01 at
25 C.

• When the "pH" symbol stops flashing,

press CFM to confirm the calibration.

The instrument is now calibrated and will remain calibrated even

CAL

HI 7004

CFM

4 cm

(1½

")

18

°C

°C

or

°C

HI 7007

RANGE

ON/OFF

when it is turned off.

Note: The meter will lose the calibration if

the batteries are removed.

HI 9023C only:
To calibrate with

manual temperature com-

pensation, follow this procedure:
• Disconnect the temperature probe and

switch the instrument ON.

• Press RANGE to select the temperature

mode.

• Place the pH electrode into the 7.01 buffer

stir gently and wait for the LCD to stabilize.

• Record the buffer temperature by using a Checktemp C or an

accurate thermometer (e.g. 18 C).

• Press the UP and DOWN to adjust the temperature accordingly.
• Press RANGE to display pH and then press the CAL key. The

value of the buffer will be displayed.

• Wait for the "pH" symbol to stop flash-

ing.

• Press CFM. The "E5" symbol will flash

to indicate that the electrode is still in
the 7.01 solution.

RANGE

CAL

and then

°C

CFM

HI 7004

Advertising