Hanna Instruments HI 9023 User Manual

Page 4

Advertising
background image

7

1) BNC electrode connector
2) Temperature probe socket
3) Liquid Cristal Display
4)

RANGE key to select pH or mV (HI 9025C only)

5)

ON/OFF key to turn the meter on or off

6)

CAL key to enter or exit calibration mode

7)

CFM key to confirm calibration

8)

Ï

°C and

Ð

C

C

C

C

C keys for manual temperature setting, or

selecting pH buffer value

9)

MEM key to store pH value in memory

10)

MR key to recall the stored value from memory

11) User-friendly graphic symbols
12) Primary display
13) Secondary display

FUNCTIONAL DESCRIPTION

FUNCTIONAL DESCRIPTION

FUNCTIONAL DESCRIPTION

FUNCTIONAL DESCRIPTION

FUNCTIONAL DESCRIPTION

HI 9024C & HI 9025C

HI 9024C & HI 9025C

HI 9024C & HI 9025C

HI 9024C & HI 9025C

HI 9024C & HI 9025C

8

SPECIFICATIONS HI 9024C & HI 9025C

SPECIFICATIONS HI 9024C & HI 9025C

SPECIFICATIONS HI 9024C & HI 9025C

SPECIFICATIONS HI 9024C & HI 9025C

SPECIFICATIONS HI 9024C & HI 9025C

HI 9024C

HI 9025C

Range

0.00 to 14.00 pH; 0.0 to 100.0 C

0.0/±399.9 mV (ISE)

±400/±1999 mV (ORP)

Resolution

0.01 pH; 0.1 C

0.1 mV (ISE)

1 mV (ORP)

Accuracy

±0.01 pH; ±0.5 C

(@20°C/68°F)

±0.2 mV (ISE)

Typical EMC

±0.01 pH; ±0.1 C

Deviation

±0.3 mV (ISE)

Calibration

Automatic with 5 memorized standard

buffers (4.01, 6.86, 7.01, 9.18, 10.01)

Offset Calibration

±1 pH

Slope Calibration

From 70 to 108%

Temperature

Automatic or manual from 0 to 100 C

Compensation

(32 to 212 F)

Electrode

HI 1230B pH electrode (included)

Temperature Probe

HI 7669/2W (included)

Input Impedance

10

12

Ohm

Battery Type & Life

4x1.5V, AA size (alkaline batteries)

approx. 400 hours of continuous use

auto-shut off after 10 minutes of non-use

Environment

0 to 50 C (32 to 122 F); 100% RH

Dimensions

meter: 196x80x60 mm (7.7x3.1x2.4")

kit: 340x230x80 mm (13.8x9x3.1")

Weight

meter: 425 g (15 oz.)

kit: 1.3 Kg (3.0 lb.)

Advertising