Validation procedure – Hanna Instruments HI 96702 User Manual

Page 15

Advertising
background image

15

VALIDATION PROCEDURE

Use the validation procedure to ensure that the instrument is properly calibrated.
Warning:

Do not validate the instrument with any standard solutions other than the HANNA CAL

CHECK™

Standards, otherwise erroneous results will be obtained.

• Turn the meter on by pressing ON/OFF.

• When the beeper sounds briefly and the

LCD displays dashes, the meter is ready.

• Place the CAL CHECK™ Standard

HI 96702-11

Cuvette A into the cuvette

holder and ensure that the notch on the
cap is positioned securely into the groove.

• Press ZERO/CFM and the lamp, cuvette

and detector icons will appear on the
display, depending on the measurement
phase.

• After a few seconds, the display will

show “-0.0-”. The meter is now zeroed
and ready for validation.