Hanna Instruments HI 96702 User Manual

Page 18

Advertising
background image

18

• After a few seconds the display will

show “-0.0-”. The meter is now zeroed
and ready for calibration. The blinking
“READ”

asks for reading calibration

standard.

• Remove the cuvette.

• Place the CAL CHECK™ Standard

HI 96702-11

Cuvette B into the cuvette

holder and ensure that the notch on the
cap is positioned securely into the groove.

• Press READ/ /TIMER and the lamp,

cuvette and detector icons will appear on
the display, depending on the
measurement phase.

• After measurement the instrument will

show for three seconds the CAL CHECK™
Standard value.

Note:

If the display shows “STD HIGH”, the
standard value was too high. If the
display shows “STD LOW”, the
standard value was too low. Verify
that both CAL CHECK™ Standards
HI 96702-11

Cuvettes, A and B are

free from fingerprints or dirt and that
they are inserted correctly.

• Then the date of the last calibration (e.g.:

”01.08.2009”

) appears on the display,

or “01.01.2009” if the factory calibration
was selected before. In both cases the
year number is blinking, ready for date
input.