ERRORS AND WARNINGS

The instrument shows clear messages when erroneous condition appears. Messages are also displayed
when the obtained values are outside expected range. The beeper is playing a beep on errors.

a) on zero reading

Light High:

There is too much light to perform a measurement.

Please check the preparation of the zero cuvette.

Light Low:

There is not enough light to perform a measurement.

Please check the preparation of the zero cuvette.

No Light:

The instrument cannot adjust the light level. Please

check that the sample does not contain any debris.

Inverted cuvettes:

The sample and the zero cuvette are inverted.

Zero:

A zero reading was not taken. Follow the instructions of the

measurement procedure for zeroing the meter.

b) on sample reading