Hanna Instruments HI 83200C User Manual
Page 34

66
67
IODINE
SPECIFICATIONS
Range
0.0 to 12.5 mg/L
Resolution
0.1 mg/L
Accuracy
±0.1 mg/L ±5% of reading at 25 °C
Typical EMC
±0.1 mg/L
Deviation
Light Source
Tungsten lamp with narrow band interference filter @ 525 nm
Method
Adaptation of the
Standard Methods for the Examination of Water and Wastewater,
18
th
edition, DPD method. The reaction between iodine and the reagent causes a pink
tint in the sample.
REQUIRED REAGENTS
Code
Description
Quantity
HI 93718-0
DPD Reagent
1 packet
REAGENT SETS
HI 93718-01 Reagents for 100 tests
HI 93718-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the
Iodine method using the procedure described in
the
Method Selection section (see page 12).
• Fill the cuvette with 10 mL of unreacted sample (up to
the mark) and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the
Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
• Remove the cap and add the content of one packet of HI 93718-0
DPD reagent. Replace the cap and shake gently for about
30 seconds to dissolve most of the reagent.
Iodine
• Add 12 drops of the HI 93704-0 reagent to
each cuvette. Replace the caps and shake
gently to mix (about 30 seconds).
• Place the blank (#1) into the holder and close the lid.
• Press
Timer and the display will show the countdown prior to zeroing the blank. The display will
show “-0.0-” when the meter is zeroed and ready for measurement.
• Remove the blank.
• Insert the cuvette with the reacted sample (# 2) into the
instrument and close the lid.
• Press
Read to start the reading. The instrument displays concentration in
µµµµµ
g/L of hydrazine.
Hydrazine
INTERFERENCES
Interference may be caused by:
Highly colored samples
Highly turbid samples
Aromatic amines
# 1
# 2
# 1
# 2
10 mL