• Remove the cuvette.

• Using the other 1 mL syringe, add to the cuvette exactly 0.5 mL of sample.

Note: Do not mix up the two syringes!

• Replace the cap and invert several times to mix.

• Reinsert the cuvette into the instrument.

• Press

Timer and the display will show the countdown prior to the measurement or, alternatively, wait

for 15 seconds and press

Read. When the timer ends the meter will perform the reading.

• The instrument displays the results in

mg/L of Magnesium (Mg).

INTERFERENCES

Interferences may be caused by:

Acidity (as CaCO

3

) above 1000 mg/L

Alkalinity (as CaCO

3

) above 1000 mg/L

Calcium (Ca

2+

) above 200 mg/L

Iron must be absent

Aluminum must be absent
Copper must be absent

Magnesium

Magnesium

SPECIFICATIONS

Range

0 to 150 mg/L

Resolution

5 mg/L

Accuracy

±5 mg/L ±3% of reading at 25 °C

Light Source

Tungsten lamp with narrow band interference filter @ 466 nm

Method

Adaptation of the Calmagite method

.

REQUIRED REAGENTS

Code

Description

Quantity

HI 93752

A-0 Mg Magnesium Buffer Reagent 1 mL

HI 93752

B-0 Mg Magnesium Indicator Reagent 9 mL

REAGENT SETS

HI 937520-01

Reagents for 50 tests

HI 937520-03

Reagents for 150 tests

For other accessories see page 128.

MEASUREMENT PROCEDURE

Note: for sample preparation follow the COLORED OR TURBID SAMPLES

procedure on page 17.

• Select the

Magnesium method using the procedure described in the

Method Selection section (see page 12).

• Using one 1 mL syringe add exactly 1.00 mL of HI 93752A-0

Mg Buffer reagent to the cuvette and use the pipette to fill the
cuvette up to the 10 mL mark with the HI 93752B-0 Mg
Indicator reagent.

• Replace the cap and invert several times to mix.

• Place the cuvette into the holder and close the lid.

• Press the

Zero key. The display will show “-0.0-” when the meter is zeroed and ready for

measurement.

MAGNESIUM

0.5 mL

of sample