Hanna Instruments HI 83200C User Manual

Page 62

Advertising
background image

122

123

Zinc

• Place the cap and insert the cuvette into the instrument

and close the lid.

• Press the

Zero key. The display will show “-0.0-” when

the meter is zeroed and ready for measurement.

• Remove the cuvette and add 0.5 mL of HI 93731B-0

Cyclohexanone to the cuvette.

Note: To prevent any contamination from the polycarbonate
cap, prior to replacing it, close the sample cuvette with
the supplied HDPE plastic stopper.

• Replace the cap and mix the sample for 15 seconds.

• Insert the sample into the instrument.

• Press

Timer and the display will show the countdown prior to the measurement or, alternatively, wait

for 3 minutes and 30 seconds and press

Read. When the timer ends the meter will perform the

reading.

• The instrument displays the results in

mg/L of zinc.

Zinc

ZINC

SPECIFICATIONS

Range

0.00 to 3.00 mg/L

Resolution

0.01 mg/L

Accuracy

±0.03 mg/L ±3% of reading at 25 °C

Typical EMC

±0.01 mg/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 575 nm

Method

Adaptation of the

Standard Methods for the Examination of Water and Wastewater,

18

th

edition, Zincon method. The reaction between zinc and the reagents causes an

orange to a dark violet tint in the sample.

REQUIRED REAGENT

Code

Description

Quantity

HI 93731

A-0

Zinc Reagent

1 packet

HI 93731

B-0

Cyclohexanone

0.5 mL

REAGENT SETS

HI 93731-01 Reagents for 100 tests
HI 93731-03 Reagents for 300 tests
For other accessories see page 128.

MEASUREMENT PROCEDURE

• Select the

Zinc method using the procedure described in

the

Method Selection section (see page 12).

• Fill one graduated mixing cylinder up to the 20 mL mark

with the sample.

• Add the content of one packet of HI 93731A-0 Zinc

reagent, close the cylinder and invert several times to mix
until completely dissolved.

• Fill one cuvette with 10 mL of the reacted sample up to

the mark.

20 mL

10 mL