Hanna Instruments HI 98280 User Manual

Page 32

Advertising
background image

32

• Press <Confirm> to accept the calibration point or <Buffer> to select an-

other buffer using the arrow keys.

• After the first calibration point is confirmed, immerse the probe in the second

buffer solution and stir gently.

• When the reading is stable and close to the selected buffer, the display shows

the “Ready” message.

• Press <Confirm> to accept the point or <Buffer> to change the buffer.
• After the second calibration point is confirmed, immerse the probe in the third

buffer solution, stir gently and wait for stable reading.

• When the calibration is completed, the display shows the following messages:

“Storing data on probe, please wait...”, “Updating GLP data, please wait ...”
and “Calibration completed”.

• Press <OK> to return to the “Calibration” menu.
• To return to the main menu, press ESC repeatedly.

Note

Note

Note

Note

Note

The pH calibration mode can be quitted at any time, by pressing the
ESC key.

Custom buffer calibration
HI9828 also allows a single point procedure to
calibrate with a custom buffer value.
• To select this option press first <Buffer> and

then <Custom> keys while the meter is wait-
ing for stable reading.

• A text box window will appear for inserting the desired custom value. The valid

range for custom buffer is from 0.00 to 14.00 pH.