• Turn the instrument ON using the power switch from

the rear panel of the instrument.

• Fill the 50 mL beaker up to the 50 mL mark with

Buffer Solution 1 (7.00 pH) or Buffer Solution 2
(8.20 pH).

• Place the pH electrode into solution and press the

CAL button. The CAL message will appear blinking
along with pH value.

• Press START STOP in order to start the electrode

calibration.

• At the end of the calibration procedure done appears

for a few seconds and then the meter automatically
returns to measurement mode.

pH CALIBRATION PROCEDURE

PUMP CALIBRATION PROCEDURE

The calibration of the pump must be performed each time the pump tube, the reagent bottle or the pH
electrode is changed. It is recommended to perform the pump calibration before each set of measurements.
• To prepare the sample for calibration, follow the

measurement procedure for Total Titratable Acidity
measurements by using HI 84102-55 Total Titratable
Acidity Standard instead of wine sample.

• After sample preparation, press the CAL PUMP button.

Std will blink on the screen.

The pH calibration must be performed each time the pH electrode is changed.