5

Range

0.0 to 25.0 g/L of tartaric acid

Resolution

0.1 g/L

Accuracy

5% of reading

Method

Acid-base titration method

Principle

End-point titration

pH Calibration

One-point in selected end-point: 7.00 pH or 8.20 pH

Sample volume

2 mL

Temperature

Automatic from 0.0 to 100.0 °C

Compensation
pH Electrode

HI 1048B (included)

Temperature Probe

HI 7662-T (included)

Pump debit

0.5 mL/min

Stirring speed

1500 rpm

Environment

0 to 50 °C (32 to 122 °F); max 95% RH non-condensing

Power supply

220V/50Hz; 10VA

Dimensions

208 x 214 x 163 mm (8.2 x 8.4 x 6.4") (with beaker)

Weight

2200 g (77.6 oz.)

REQUIRED REAGENTS
Code

Description

Quantity/test

HI 84102-50

Titrant

HI 84102-55

Standard

2 mL

HI 84102-60

Buffer Solution 1 pH 7.00

50 mL

HI 84100-61

Buffer Solution 2 pH 8.20

50 mL