Hanna Instruments HI 9146 User Manual

Page 12

Advertising
background image

12

A

AA

AA u t o E n d

u t o E n d

u t o E n d

u t o E n d

u t o E n d

To freeze the first stable reading on the LCD
press

AutoEnd while the instrument is in

measurement mode.
The ”

HOLD” tag will be displayed blinking on

the LCD until the reading will stabilize.
When the reading is stable, the “

HOLD” tag

stops blinking and the reading is frozen on
the LCD.
Press

AutoEnd again to return to normal

measurement mode.

Note: • Pressing Range the instrument will skip to the displayed

range, without leaving AutoEnd mode.

• Pressing

Setup then GLP, the instrument leaves AutoEnd

mode and performs the selected function.

HI 7040

HI 7040

Calibrate the instrument frequently, especially if high accuracy is required.
The instrument can be calibrated in maximum 2 points: 0.0% (

zero

calibration) and 100.0% (slope calibration).
The calibration of this instrument is very simple.
Before proceeding with the calibration, make sure the probe is ready for
measurements (see page 8), i.e. the membrane cap is filled with electrolyte
and the probe is connected to the meter and properly polarized.
For an accurate calibration, it is recommended to wait at least
15 minutes to ensure precise conditioning of the probe.
The zero calibration of the

HI 9146 is very stable,

therefore this procedure needs to be performed only
whenever the probe or the membrane is replaced.
However, because the slope calibration is more critical,
it is recommended to perform this procedure every week.

D O C A L I B R A T I O N

D O C A L I B R A T I O N

D O C A L I B R A T I O N

D O C A L I B R A T I O N

D O C A L I B R A T I O N

Advertising