Hanna Instruments HI 9146 User Manual

Page 13

Advertising
background image

13

INITIAL PREPARATION

• Pour small quantities of

HI 7040 Zero Oxygen solution into a

beaker. If possible, use a plastic beaker to minimize any EMC
interferences.

• Make sure the probe is ready for measurements (see probe preparation

on page 8), i.e. the membrane is filled with electrolyte and the
probe is connected to the meter.

• Switch the meter on by pressing the

On/Off switch.

• For an accurate calibration, it is recommended to wait for at least

15 minutes to ensure precise conditioning of the probe.

• Remove the protective cap from the DO probe.
• Set the appropriate altitude factor (see

page 19). Make sure the salinity factor is
set to zero (see page 18).

ZERO CALIBRATION

• Submerge the probe into

HI 7040

zero oxygen solution and stir gently
for 2-3 minutes.

• Press

CAL. The “ ” and “NOT READY”

tags will blink on the LCD until the
reading is stable.

• When the reading is stable,and is within

the limits (±15% f.s.) “

CFM” starts

blinking. Press

CFM to confirm the

0.0%” DO calibration.

• Press

CAL. The instrument will return to measurement mode and will

memorize the zero calibration data. For a
two-point calibration do not press

CAL and

follow the procedure below.