Hanna Instruments HI 9146 User Manual

Page 8

Advertising
background image

8

INITIAL PREPARATION

The instrument is supplied complete with batteries. In order to place the
batteries inside the instrument follow the instructions from page 21.
To take measurements, connect the DO probe to the meter securely by
aligning the pins with the socket located on the back of the meter,
pushing the plug in and tightening the threaded ring.
Probes shipped from Hanna Instruments are dry. To hydrate the probe
and prepare it for use, connect it to the meter and proceed as follows:

1. Remove the red and black plastic cap.

This cap is for shipping purposes and
can be thrown away.

2. Wet the sensor by soaking the bottom

2½ cm (1") of the probe in electrolyte
(

HI 7041S) for 5 minutes.

3. Rinse the membrane cap (

HI 76407A

supplied in the kit with the meter)
with electrolyte solution while shaking
it gently. Refill with clean electrolyte
solution.

4. Tap gently the sides of the membrane

cap with your finger tip to ensure that
no air bubbles are trapped. To avoid
damaging the membrane, do not tap
it directly on the bottom.

5. Make sure that the rubber O-ring sits

properly inside the membrane cap.

6. With the sensor facing down, slowly

screw the cap clockwise. Some electrolyte
will overflow.

When not in use and during polarization
(see page 9), use the protective transparent
cap supplied in the kit with the meter.

Shipping

cap

black

red

FILL FIRST

THEN TAP

THEN SCREW

BACK ON

O P E R A T I O N A L G U I D E

O P E R A T I O N A L G U I D E

O P E R A T I O N A L G U I D E

O P E R A T I O N A L G U I D E

O P E R A T I O N A L G U I D E