Hanna Instruments HI 9141 User Manual

Page 13

Advertising
background image

25

24

FUNCTION

UP

DOWN

ALTITUDE COMPENSATION

Press the FUNCTION key twice and “F1” will be
displayed.

Use the UP and DOWN keys to set the altitude
between 1 and 1900 m, in steps of 100 m
(1 meter = 3.28 feet).

Altitude affects the disssolved oxygen concen-
tration by decreasing its value. The following
table reports the maximum oxygen solubility at
various temperatures and altitudes.

FUNCTION

UP

DOWN

35 g/L

9.0
8.6
8.3
8.0
7.6
7.4
7.1
6.9
6.6
6.4

30 g/L

9.3
8.9
8.6
8.2
7.9
7.6
7.3
7.1
6.8

6.6

°C

10
12
14
16
18
20
22
24
26
28

0 g/L

11.3
10.8
10.3

9.9
9.5
9.1
8.7
8.4
8.1
7.8

10 g/L

10.6
10.1

9.7
9.3
8.9
8.5
8.2
7.9
7.6
7.4

20 g/L

9.9
9.5
9.1
8.7
8.4
8.0
7.8
7.5
7.2
7.0

Salinity (g/L) at Sea Level

°F

50.0
53.6
57.2
60.8
64.4
68.0
71.6
75.2
78.8
82.4

°C

0
2
4
6
8

10
12
14
16
18
20
22
24
26
28
30
32
34
36
38
40

0 m

14.6
13.8
13.1
12.4
11.8
11.3
10.8
10.3

9.9
9.5
9.1
8.7
8.4
8.1
7.8
7.5
7.3
7.1
6.8
6.6
6.4

300 m

14.1
13.3
12.7
12.0
11.4
10.9
10.4

9.9
9.7
9.2
8.8
8.4
8.1
7.8
7.5
7.2
7.1
6.9
6.6
6.4
6.2

600 m

13.6
12.9
12.2
11.6
11.0
10.5
10.1

9.6
9.2
8.7
8.5
8.1
7.8
7.5
7.3
7.0
6.8
6.6

6.3

6.2
6.0

1200 m

12.7
12.0
11.4
10.8
10.3

9.8
9.4
9.0
8.6
8.3
7.9
7.7
7.3
7.0
6.8
6.5
6.4
6.2
5.9
5.7
5.6

1500 m

12.3
11.6
11.0
10.4

9.9
9.5
9.1
8.7
8.3
8.0
7.7
7.3
7.1
6.8
6.6
6.3
6.1
6.0
5.7
5.6
5.4

1800 m

11.8
11.2
10.6
10.1

9.6
9.2
8.8
8.3
8.0
7.7
7.4
7.1
6.8
6.6

6.3
6.1
5.9
5.8
5.5
5.4
5.2

°F

32.0
35.6
39.2
42.8
46.4
50.0
53.6
57.2
60.8
64.4
68.0
71.6
75.2
78.8
82.4
86.0
89.6
93.2
96.8

100.4
104.4

900 m

13.2
12.4
11.9
11.2
10.6
10.2

9.7
9.3
8.9
8.6
8.2
7.8
7.5
7.3
7.0
6.8
6.6
6.4
6.1
5.9
5.8

Altitude, Meters above Sea Level

TEMPERATURE COMPENSATION

The DO probe features a built-in temperature
sensor, so that DO readings are automati-
cally compensated for temperature effects.

SALINITY COMPENSATION

Press FUNCTION key until “F2” is displayed.

Use the UP and DOWN keys to set the salinity
between 0 and 40 g/L.

Salinity affects the dissolved oxygen concen-
tration by decreasing its value. The table below
shows the maximum solubility of oxygen at
various temperature and salinity.

Advertising
This manual is related to the following products: