Approximate run times for the following measurements are shown below:

HG (without

7-9 min.

SL(without B2)

9 min

DS

3 min

ZS

4 min

UV

8 min

UX

6 min

UM

Indeterminate must be aborted with
HOME key or new schedule entry

FM

11 min


Writing Schedules:

•

Determine the Scientific Objectives, and draw up a table of Solar Zenith Angle vs.
Commands.

•

Use the SA command to assist in relating to local time.

•

From the Brewer main menu, send the Command SE.

•

a table will appear on the screen with two columns, one for SZA, and one for the commands
to be executed at the SZA.

•

starting with the first SZA after solar midnight (it will probably be -ve), enter the SZA in the
left hand column of the table, and enter the commands to be executed at that angle in the
right hand column.

•

carry on through the day, remembering that the SZA changes to +VE after noon (no sign
defaults to +ve).

•

when the schedule is finished, press Ctrl+End and follow the instructions for saving the
schedule.

•

the schedule can be tested using the SK or SKC routines.