Kipp&Zonen CNR 4 Net Radiometers User Manual

Page 19

Advertising
background image





18

2.2.1

Specifications of the pyranometer

Pyranometers Specification

Unit

Value (All indicated values are absolute values)

Spectral range

nm

300 - 2800 (50% points)

Definition

Instrument sensitivity within a specific spectral range

Sensitivity

µV/ W/m²

10 to 20

Definition

Calibration factor

Impedance

Ω

20 to 200, typically 50

Definition

Typical resistance measured at the output

Response time

s

< 18 (95% response)

Definition

Sensor response time

Non-linearity

%

< 1 (from 0 to 1000 W/m

2

irradiance)

Definition

Maximum deviation from the responsivity at 500 W/m

2

due to any

change of irradiance within the indicated range.

Temperature dependence of
sensitivity

%

N/S (- 40 °C to -10°C)

< 4 (- 10 °C to +40 °C)

N/S (+40 ºC to +80 ºC)

Definition

Maximum error due to any change of instrument temperature within
the indicated temperature interval.

Tilt error

%

< 1

Definition

Maximum deviation from the responsivity when tilted at any angle
and at 1000 W/m² irradiance .

Zero offset A

W/m²

< 15 (0 to -200 W/m² / IR net irradiance)

Definition

Caused by cooling of the dome due to sky radiation

Zero offset B

W/m²

< 3 (at 5 K/h temp. change)
< 1 (with CVF 4 installed)

Definition

Response to change in ambient temperature.

Field of view:
Upper detector
Lower detector


180º
150º (due to lower sun shield. To prevent illumination at low zenith
angles)

Definition

Sensor opening angle

Directional error

W/m²

< 20 (angles up to 80° with 1000 W/m² beam radiation)

Definition

Combined zenith and azimuth error from 0°- 80

°

with 1000 W/m²

beam radiation