Cnr 4 net radiometer, Temperature – Kipp&Zonen CNR 4 Net Radiometers User Manual

Page 8

Advertising
background image





7

1.1.3

The radiometer Temperature connector (T)

The temperature connector of the CNR 4 carries the signals for the PT-100 and Thermistor. The PT-100 and
Thermistor have identical accuracy, the reason for selecting one or the other is mainly the data logger involved. The
Pt-100 temperature sensor has 4 wires, two for the measuring current and two for measuring the voltage over the
resistor (100 Ohm @ 0 °C) . In this way the measurement accuracy is minimally affected by the cable length. The
thermistor has a higher impedance of 10 kOhm and is therefore less influenced by cable length but has no
compensation for it. The shield of the cable is twisted together and covered with a black sleeve. If the CNR 4 (rod)
is grounded to the mast it is advised not to connect the shielding on the data logger side. This might cause ground
loops and offsets in the signal.

Figure 1.3 The temperature connections of the CNR 4. The instrument has one (4-wire) Pt-100 output and a
Thermistor with a 2-wire connection.

CNR 4 NET RADIOMETER

TEMPERATURE

6 WIRE CABLE • 6-ADRIGES KABEL • CÂBLE 6 FILS • CABLE DE 6 CONDUCTORES

7

8

5

3

4

6

Grey

Grau • Gris • Gris

Combined

Kombiniert

Combiné

Combinado

Pt-100

Standard

Standard

Etalon

Estándar

Combined

Kombiniert

Combiné

Combinado

Brown

Braun • Brun • Marrón

Green

Grün • Vert • Verde

Yellow

Gelb • Jaune • Amarillo

White

Weiss • Blanc • Blanco

Thermistor

Standard

Standard

Etalon

Estándar

Black

Schwarz • Noir • Negro

1

2

3 4

5

6

7

8

Ground *

Erde

Terre

Tierra

Housing

Gehäuse

Boîte

Cubierta

Shield

Abschirmung

Protection

Malla

Advertising