General Installation Notes

The Two and Six-Place Kjeldahl Digestion Systems are shipped fully assembled.
The Six-Place Kjeldahl Digestion System will require exhaust duct work which
should be made by a licensed contractor. The Two-Place Kjeldahl Digestions
System can be installed by the customer inside a fume hood and the fume removal
manifold connected to a common laboratory faucet water aspirator. The customer
will need to purchase the required Kjeldahl flasks,

Use caution when you remove the Six-Place Kjeldahl Digestion system protective
packaging material taped to the heater runways as fragile components are
involved. Levelers are shipped installed.

Location

The Two-Place Kjeldahl Digestion System can be carried by hand to its final
location inside a fume hood.

The Six-Place Kjeldahl Digestion System is heavy and cannot be disassembled to
reduce the overall size. Special arrangements should be made in advance for
moving the system to its final location. Special instruction tags are attached to the
apparatus; they must not be removed until the installation has been completed.

The Six-Place Kjeldahl Digestion Systems do not include a surrounding safety
enclosure. It is strongly advised to install this product in a segregated location
away from normal laboratory personnel traffic and inside a floor mounted
chemical fume hood. High heat and boiling acid present a safety hazard to
personnel in close proximity to the Six-Place Kjeldahl Digestion System and the
Six-Place Kjeldahl Digestion System should be placed inside a floor mounted
fume hood. Also recommended is spot exhaust ventilation for the final
installation location to remove excess heat produced by the digestion heaters.