background image

Chapter 3: Normal Operation & Routine Maintenance

Product Service 1-800-522-7658

11

Clean-Up and Cosmetic Guidelines

• Keeping the Kjeldahl Digestion System clean will preserve the

appearance and improve life of the equipment.

• The equipment can be kept clean by washing with a weak solution of

sodium hydroxide and rinsing with clear water.

• Sulphate, a salt of Sulfuric Acid, may build up between the channel

support and the fume removal manifold. Neutralize this with a weak
solution of sodium hydroxide periodically to keep the unit clean.

Fume Removal Manifold Suction Adjustment

The suction along the fume removal manifold is adjusted as follows:

Two Place Digestion System
Suction is adjusted by changes in the water supply flow rate. The suction
may be decreased by decreasing the flow rate.

Blower Exhaust System
No adjustment is required on the suction pressure at the manifold when used.
The remote blower should be inspected annually for debris, proper operation,
and checked for adequate airflow as specified on pages 11-12 of this manual.

Exhaust Systems Maintenance

Sometimes the suction in the nipples is reduced due to obstruction within the
exhaust system. These obstructions can usually be washed from the fume
pipe, piping assembly or water ejector by boiling water in a number of flasks
placed on the digestion heaters. The blower or water aspirator should be in
operation to draw the steam through the fume pipe.

Digestion Drip Shield

The drip shield is constructed of stainless steel and is resistant to most acid attack.
Acid spills can however cause discoloration and eventual marring of the surface
unless the following proper procedures are followed in their clean up.

1. Promptly wash down all major acid spills contained by the shield.
2. Periodically clean the stainless surface with any residential

stainless steel sink cleaning compound.

3. After cleaning, renew the bright shiny appearance by lightly

sanding with 220 grade emery paper.

The Two-Place Kjeldahl Digestion Systems does not feature a drip shield
with a drained tube.