Chapter 2: Installation

Product Service 1-800-522-7658

7

Electrical

The Two-Place Kjeldahl Digestion System comes with an 8-foot, 3 wire cord and
plug and can be connected to a 115 volt 60Hz 11 Amp circuit.

The Six-Place Digestion Kjeldahl System has been wired at the factory per the
product model number ordered by the customer. Labconco model numbers ending
in -11 are 115 volts. Model numbers ending in -12 are 230 volts. Models ending
in -13 are 230 volt, three phase. Amperage requirements vary depending on the
customer selected operational voltage of the digestion rack. The voltage and
amperage tables below provide the electrical requirements for the Six-Place
Digestion Systems. A qualified electrician will need to connect the facility
electrical supply to the main breaker box of the Digestion System.

1. Remove circuit breaker box cover panel.
2. Main line lead connection terminals are identified and connections

must be made accordingly.

3. Line leads to the apparatus must conform to local electrical codes.
4. The Remote Blower can be wired to the on/off switch to control the

remote blower from the unit. Remote Blower MUST be on before
using the unit.

5. Provide an electrical ground to the apparatus per code.
6. Before applying power to apparatus, check the electrical panel and

breakers for loose connections.

7. Reinstall box cover panel.
8. Power the breaker box and reset breakers to check circuits in the

apparatus.

9. All electrical wiring and connections must conform to local codes and

should be performed by qualified electricians. An earth ground must
be provided.



P/N AC Electrical Code

Amps

-11 115v, single phase, 60 Hz

33

-12 230v, single phase, 60 Hz

18

-13 208/230v, three phase, 60 Hz

10