Chapter 3: Getting Started

Product Service 1-800-522-7658

7

Original instructions


Plus the following:

Part #

Component Description

7314000 User’s

Manual

7396206 Coupling

Insert

7455100

Rotor, 12-13 mm (Standard Models)

7455101

Rotor, 12-13 mm PTFE (Acid Resistant Models)

7450200 Glass

Jar


If you do not receive one or more of the components listed for your CentriVap, or
if any of the components are damaged, contact Labconco Corporation
immediately for further instructions.

Setting Up Your CentriVap

After you verify receipt of the proper components, move your CentriVap to the
level and sturdy location where you want to install it. There is no need to secure
the CentriVap to the work surface. Then, follow the steps listed below.

Emergency Access Into the Chamber

The CentriVap is designed to prevent access to the chamber in the event of a
power disruption. If it is necessary to open the lid when there is no electrical
power connected to the CentriVap, insert a small screwdriver or similar
instrument into the small round hole on the left side of the case near the top
behind the control panel. This will unlock the lid latch mechanism. While
holding the screwdriver in place, raise the lid with the other hand.

CAUTION: Never attempt to defeat the latch or open the lid while the
CentriVap is running. Personnel injury can result from moving parts and
chemicals.