Chapter 7: Troubleshooting

Product Service 1-800-522-7658

25

Original instructions


Other corrective actions for potential problems are as follows:

PROBLEM

CAUSE

CORRECTIVE ACTION

Unit will not operate

Unit not connected to
electrical power

Connect unit to proper electrical
receptacle.

Circuit breaker blown

Correct electrical problem and reset circuit
breaker by pressing button.

Lid open

Close lid.

Excessive vibration

Sample tubes not located
symmetrically in rotor

Reposition sample tubes.

Sample odor in lab

Vent hose exhausting into
lab area

Redirect hose to fume hood.

Evaporation rate is
reduced

Heater inoperable

Contact Labconco.

Vacuum pump failure

Check pump.

Obstruction in hose

Remove obstruction or replace hose.

Lack of adequate vacuum See below.

No vacuum/poor vacuum

Pump not on

Turn on pump.

Control valve open

Check control valve.

Leaks in lines or
connectors or gasket

Locate and repair.

Foreign material on lid
gasket

Clean gasket and lid.

Pump is not functioning
properly

Check pump by locating vacuum gauge
closer to pump and close off rest of system.

If pump is faulty, seek authorized service
or replace pump.

Cold Trap or
Concentrator gasket is
not sealing properly

Check gasket for cleanliness. Adjust
gasket as needed.

Cold Trap lid not seated

Hold lid down until vacuum is initiated.

Ice formed on Cold
Trap lid sealing surface

Defrost and wipe dry.