INSTALLATION

10

Location

The Gel Dryer should be placed on a stable level surface, not exposed to extremes in
temperature or high humidity.


Electrical Connection

The Gel Dryer requires a dedicated electrical outlet rated to meet the requirements of
both the Gel Dryer and the vacuum pump. The Gel Dryer rated for 115 VAC operation
requires a 20 amp power source, while the 230 VAC model requires 10 amp service.