Labconco Gel Dryer 4330150 User Manual

Page 8

Advertising
background image

INTRODUCTION

8

Component Identification (115V unit shown)
1. Gel Dryer Fuse. The Gel Dryer Fuse protects the unit from an electrical overload.

The fuse is rated at 10 amps time delay (2 fuses, each 10 amps time delay for the 230
model).

2. Platen. The cast aluminum platen allows for a uniform distribution of heat during

drying. The platen is Teflon® coated

to minimize sticking.

3. Screen. The porous screen supports the gel during the drying process, preventing it

from sticking to the platen.

4. Porous Polyethylene Sheet. The porous polyethylene sheet covers the gel during the

drying process, preventing the gel from sticking to the silicone rubber sheet.

5. Temperature Adjustment Knob. The drying temperature is controlled by setting the

adjustment from 40° to 80°C.

6. Heater Timer. The Heater Timer sets the length of time that the heater operates. It

may be adjusted from 0 to 4 hours (0 to 4 hours on the 230 volt model).

7. Heater Indicator Lamp. The Heater Indicator Lamp illuminates while the heating

element is in operation. As the platen temperature reaches the selected drying tem-
perature, the light will begin to blink, and then go out when the preset temperature is
reached.

8. Vacuum Inlet. The l/4 inch hose barb on the rear of the platen connects the Gel

Dryer to the vacuum source.

9. Vacuum Pump Timer. The Vacuum Pump Timer sets the length of time that the

separate vacuum pump (not included) operates. It may be adjusted from 0 to 4 hours
(0 to 4 hours on the 230 volt model).

10. Vacuum Pump Electrical Outlet. The 10 amp outlet (5 amp for the 230 volt model)

is directly controlled by the Vacuum Timer.

11. Vacuum Pump Outlet Fuse (115V units). The Vacuum Pump Outlet Fuse protects

the Vacuum Pump Electrical Outlet from an overload. The fuse is rated at 10 Amps
time delay. 230V Units: There is no fuse specifically for this outlet on 230 volt units.
The fuses in Item #1 above (2 each at 10 amps time delay) protect the units entire
electrical system from overloads.

12. Silicone Rubber Sheet. The flexible rubber sheet provides a vacuum seal during the

drying process.

13. Mylar Sheet. (not shown) Is used in place of the porous polyethylene sheet.

*Teflon® is a registered trademark of E.I. Du Pont de Nemours & Company, Inc.
*Mylar® is a registered trademark of E.I. Du Pont de Nemours & Company, Inc.

Advertising
This manual is related to the following products: