TROUBLESHOOTING

20

PROBLEM

CAUSES

CORRECTIVE

ACTION


Heater and vacuum pump

Gel Dryer fuse

Replace fuse with the

controls don’t work

blown

proper

replacement


Circuit breaker

Reset circuit breaker

tripped

Heater control works but

Vacuum pump

Turn on vacuum

vacuum pump doesn’t

switch in “off”

pump switch


Vacuum pump

Replace fuse with the

fuse blown

proper replacement


Vacuum seal doesn’t

Vacuum leak

Check connection

form, or forms slowly

Insufficient vacuum

Vacuum pump should

pump capacity

have a capacity of

50

liters/minute

or

more


Support screen bent

Ensure that screen

or damaged

lays flat


Silicone rubber

Moisten the surface of

sheet and platen dry

the rubber sheet and

platen

with

water

before

applying

vacuum.


Rubber sheet/platen

Inspect the surfaces

surface damaged

and replace if

necessary


Gel cracks/shatters

Insufficient drying

Increase the drying

time

time


Insufficient vacuum

The vacuum source

should

have

a

minimum

free

air

displacement

of

50

liters/minute

and

develop

a

minimum

vacuum of 25 in. Hg.