Chapter 3: Getting Started

Product Service 1-800-522-7658

8

Emergency Access into the Chamber

The CentriVap is designed to prevent access to the chamber in the event of a
power disruption. If it is necessary to open the lid when there is no electrical
power connected to the CentriVap, insert a small screwdriver or similar
instrument into the small round hole on the left side of the case behind the control
panel. This will unlock the lid latch mechanism. While holding the screwdriver
in place, raise the lid with the other hand.

CAUTION: Never attempt to defeat the latch or open the lid while the
CentriVap is running. Personnel injury can result from moving parts and
chemicals.






Component Orientation & Hose Connections

The relative position of the
CentriVap, Cold Trap and
Vacuum Pump should be as
shown to optimize the air flows of
the refrigeration systems. There
must be at least 6 inches between
the Refrigerated CentriVap and
the Cold Trap. More space is
better. There must be at least 3
inches between the back of the
refrigerated CentriVap and the
adjacent wall surface. Restriction
of the airflow through the cabinet
during operation could adversely
affect performance.