Labconco Refrigerated CentriVap Centrifugal Concentrators 7310041 User Manual

Page 31

Advertising
background image

Chapter 7: Troubleshooting

Product Service 1-800-522-7658

27

PROBLEM

CAUSE

CORRECTIVE ACTION

Frequent oil change needed in
pump

Secondary Trap insert is spent

Change insert often.

Cold Trap is not emptied after
each run and dried

Empty the traps (glass or stainless steel) after
each run and replace.

Vacuum too strong for chemical

Use a Secondary Trap insert and diaphragm
pump.

Unit starts and shuts off

Latch optical sensor improperly
calibrated

Unplug power cord. Wait 10 seconds. Plug in
power cord.



Advertising