Chapter 2: Prerequisites

Product Service 1-800-522-7658

4

Original instructions

Exhaust Requirements

If you intend to connect the Procedure Station to the optional Ventus

TM

Canopy Connection Kit, first examine the location to ensure that it
accommodates the cabinet’s exhaust duct. The area directly above the
cabinet’s exhaust port should be clear of structural elements, water and utility
lines, or other fixed obstructions. There should be enough clearance to
accommodate a 10-inch diameter duct. Avoid cabinet locations that require
an elbow directly above the cabinet’s exhaust connection or an excessive
number of elbows in the exhaust system. For further information about the
cabinet’s exhaust system requirements, please refer to Chapter 3: Getting
Started.

Electrical Requirements

The procedure station models have the following electrical requirements:

Model #

Requirements

31x4xxx0x

115 VAC, 60 Hz, 12 Amps

31x4xxx1x

100 VAC, 50/60 Hz, 12 Amps

31x4xxx-20, 30, 40, 50, 60, 70

230 VAC, 50/60 Hz, 6 Amps

31x6xxx0x

115 VAC, 60 Hz, 16 Amps

31x6xxx1x

100 VAC, 50/60 Hz, 16 Amps

31x6xxx-20, 30, 40, 50, 60, 70

230 VAC, 50/60 Hz, 8 Amps

Note: A dedicated outlet with an appropriate circuit breaker should be
located as close as possible to the right rear side of the cabinet, at a
height even with, or higher than, the top of the bench or stand. Consult
your local electrical codes for properly rated circuit breakers. For safe
operation the dedicated outlet must provide the protective earthing
ground connection to the cabinet.

Note: On 100 and 115 VAC models, both electrical outlets are protected by a
ground fault interrupter circuit (GFIC). Labconco does not recommend
plugging the procedure station into a GFIC outlet.

Electrical outlets in the cabinet are restricted to 5 amps
maximum current.

Prises électriques dans l'armoire sont limitées à 5 courant
maximum ampères.

Table 2-1