Initial Certification

Prior to use, a qualified certifier should certify all Procedure Stations. Under
normal operating conditions, the station should be recertified at least annually
and when relocated or serviced. The certifier should perform the following
tests, as recommended in ANSI/NSF International Standard Number 49 in
effect when the cabinet was manufactured:

• Downflow Velocity Profile Test

• Inflow Velocity Test

• Airflow Smoke Patterns

• HEPA Filter Leak Test

• Optional Canopy Alarm Test and Operation

• Vibration Test *

• Noise Level Test *

• Lighting Intensity Test *

*These tests are user comfort related tests and may be omitted at the

user’s or certifier’s discretion.