5.1 Trouble Shooting

The Heartsim 200 is designed to be a reliable unit requiring a
minimum of maintenance. With this in mind, there are only a few
faults the user can rectify. In all other cases contact Laerdal
Medical, or your Laerdal dealer for authorized servicing or repair.

Problem

Correction

Battery Low LED is on

- Replace batteries

No ECG signals displayed

- Check all connections

No pulse simulation

- Check to see if battery

low LED is on.

- Check manikin cable

connection.

- Check type of manikin

(see page 4).

- ALS Skilltrainer only:

Pulse is pressed down on
both sides simultaneously.

5.2 Servicing

5.2.1 Battery Removal/Replacement
To remove and replace the batteries:
- Remove the battery cover on the rear of the Heartsim 200.
- Remove and safely discard the old batteries.
- Install the new batteries, AA size (R6) 1.5V, in the correct

orientation as shown on the cabinet bottom.

- Re-fit the battery cover.

5.3 Cleaning

5.3.1 General
Cleaning of the Heartsim 200 is restricted to general cleaning of
the keyboard and case:
- Moisten a cloth with a suitable cleaning agent. Use a mild soap

or cleaner.

- Gently wipe the surface of the keyboard and case.
- Remove any excess liquid with a dry cloth.
- Do not immerse the unit in water.

13

5 . M a i n t e n a n c e