Troubleshooting, Lit er – Laerdal LGH MamaNatalie User Manual

Page 9

Advertising
background image

17

Disassembly and reassembly

It may be practical to disassemble MamaNatalie before cleaning or other maintenance. In order to

disassemble, follow the below steps. To reassemble, follow the steps in the reverse order.

lit

er


a) Remove the Stomach Skin from the top

of the Skeleton

b) Detach the Stomach Skin from the back of

the Skeleton.

c) Pull gently on the Blood Tube to release it

from the Blood Nipple.

d) Twist the Blood Nipple 90 degrees to

release from Skeleton.

Tip: When reassembling, position the blood nipple into the opening of the Skeleton with the inlet tube
pointing downwards. Twist 90 degrees to fixate.

lit

e

lit

e

it

e

e

li

r

r

r

r

e) Release the Uterus Strap by opening the

press-stud band. The skin is now loose and

can be removed from the Skeleton.

Note: When reassembling, make sure the urethra
is positioned well into the small opening of the
Uterus Strap before closing the Uterus Strap
around the skeleton. When correctly positioned,
the urethra should be visible for the Learner from
the outside.

Transport

MamaNatalie

can be transported ready for simulation.

When transporting a water filled NeoNatalie, use the storage pouch included in the NeoNatalie kit.

Alternatively transport NeoNatalie inside the Uterus of MamaNatalie.

Troubleshooting

Artificial blood flows when it should not
- Check that the Blood Valve is in the closed position.

- Check all components for leakage.

- If the Blood Tube seems to be blocked:

a) Fill the Penguin with water and try to flush the water through the tube from the blood outlet

inside the uterus.

b) Inspect the tube from the Blood Tank. If the tube is deformed, massage the relevant point

with your fingers until the tube has opened.

Hard to push NeoNatalie’s head out of the vaginal opening
- Check that the Uterus hangs on the hook of the Blood Tank Cap.

- Check the position of the head.

- Check that the Stomach Skin is fixed with the Uterus Strap to the pubic bone.

- Try lubricating NeoNatalie’s head with a few drops of baby oil.

Urine does not work as expected
- Inspect the Urine Bladder for air. If air is present remove air by squeezing the bladder with a

urine catheter inserted into the urethra. Fill the Urine Bladder with water as instructed.

- While the Learner is catheterizing, the Operator can assist the fluid flow by gently pressing

directly onto the Urine Bladder located behind the pubic bone.