Replacing and Filling Tibial IO Module with Blood
1 Attach the Tibial IO bag to the Tibial tube and close off
the pinch clamp.
2 Roll the leg band, to expose the Tibial IO module.
3 Remove the IO tape. Then remove the Tibial IO unit
from the leg.
4 Remove the tube from the Tibial IO module.
5
Remove the Tibial IO Pad from the Tibial IO
chassis.
6 Before replacing the new Tibial IO, ensure that the
nipple is retracted in the Tibial IO Pad.
7 Fit the new Tibial IO pad into the chassis.
8 Secure the Tibial IO pad in place by pressing the rear
of the pad with thumbs until the nipple moves forward
and locks the unit in place.