Replacing blood system filter, Servicing the patient simulator, Maintenance – Laerdal SimMan 3G User Manual
Page 41: Always perform a service

MAINTENANCE
Replacing Blood System Filter
If reduced blood flow is experienced, the filter may be clogged and
needs replacement.
Never run the Patient Simulator without a filter.
Bloo
d R
igh
t L
eg
1 Turn off the Patient Simulator.
2 Remove genitalia with the catheterization assembly for easy
access.
3 Disconnect the filter from the right leg and pelvis blood tubes
and remove it.
4 Connect a new filter by reversing the steps above.
Servicing the Patient Simulator
A full service, including cleaning of the base board, should be
performed at regular intervals.
Caution: All servicing must be performed by qualified service
personnel.
Always Perform a Service:
− If liquids have been spilled in the Patient Simulator
− After use in dusty environments.
Caution: Do not use cables or connectors showing visible damage.
41