Enabling co, Simulation, Calibration – Laerdal VitalsBridge User Manual
Page 31: Vitalsbridge directions for use
Advertising

VitalsBridge Directions for Use
30
rev. B.7
PN 10229
CO
2
Calibration
Calibration may be needed if the end-tidal CO
2
reading on the patient vital signs monitor does not
properly match the CO
2
waveform on the Instructor Application. From the “Calibrate” tab on the
VitalsBridge software, select “(Update)” next to the CO2 label. Select a “(New Calibration)”.
Enter the name of the custom calibration.
For each calibration point, observe the target etCO
2
value in the CO2 Calibration window. Slowly move
the indicator to adjust the flow up or down until the end-tidal CO
2
reading on the patient vital signs
monitor equals the target end-tidal CO
2
reading.