3.6 Leveling the weighing machine
To function properly, the balance must be precisely horizontal. The balance is fitted with one bubble level. 
and adjustable feet for level-control, with the aid of which it is possible to compensate for small height 
differences and / or any unevenness in the surface on which the balance is positioned. 
 
The screw feet must be adjusted so that the air bubble is precisely in the center of the sight glass of the 
bubble level (see Fig. 1) 
 
Place the scale horizontally and keep the bubble inside the bubble level aligned with the red circle (See 
Fig. 1). In order to get exact measurements, the balance must again be carefully leveled after each 
relocation. 
 
 
Fig. 1 Correct leveling with the aid of the bubble level 
 
 
ER
3.7 Calibrating the weighing machine 
 
Since the Earth’s gravity is not the same everywhere, each balance must be adjusted to compensate for 
the gravity differences at each location in accordance with the underlying physical weighing principles. 
This adjustment process, known as calibration and must be carried out on initial installation and after 
each subsequent relocation. 
 
In order to ensure exact measurements, it is recommended that the balance should be calibrated 
regularly using a known external calibration weight (see Section 5 below).