2 program options, 1 interface settings, 2 parameter settings – LW Measurements HRB 1002TL User Manual

Page 17

Advertising
background image

17

4.2 Program options


Please see our web-site at http://lwmeasurements.com for practical demonstrations of application usage.


4.2.1 Interface Settings

Press [ON/OFF] to turn on the scale, when instrument displays --------

Press [PRINT], the display will show b xxx BAU value

Press [UNIT] to select among (1200,2400,4800,9600), default setting is 9600

Press [ZERO] to enter Parity mode, the display will show P xxx PAR

Press [UNIT] to select odd, even or none ( odd stands for 7 data bits with odd parity, even stands

for 7 data bits with even parity, none stands for 8 data bit without parity), default setting is none

Press [ZERO] to enter Print mode, the display will show n xxx mod

Press [UNIT] to select CON,OFF, KEY, STB

a)

OFF stands for serial data output disabled

b)

KEY stands for manual mode by pressing [PRINT]

c)

STB stands for automatic print when scale is stable

d)

CON stands for continuous print

Default setting is KEY

Press [ZERO] again to return to weighing mode.

4.2.2 Parameter settings

4.2.2.1 Setting zero tracking range

Press ON/OFF key, and when

“- - - - - -

“ shows, press PCS key, display will show “Zrd x.x”, press UNIT

key to select, factory default setting is 0.8


4.2.2.2 Setting zero tracking time

Press ON/OFF key , and when

“- - - - - -

“ shows, press PCS key, and then press ZERO key till display

shows

“Zrt x.x”, press UNIT key to select, factory default setting is 2.0

4.2.2.3 Setting displaying zero range

Press ON/OFF key, and when

“- - - - - -

“ shows, press PCS key, and then press ZERO key multiple times

till display shows

“rzd x.x”, press UNIT key to select, factory default setting is 0.8