5 Set-up and connection

26

Version 11/2008, Rev. 2.2 of 04/12/2013 • sb

Translation of the original operating manual

Pos: 94 / 200 Christ /37 0 RVC-BA (STANDARDM ODULE)/ 050 Au fstellu ng u nd Ans chluss/ 050 -00 30 B elüftu ngsv entil @ 8\m od_ 130 941 9388 700 _68 .docx @ 4 6708 @ 3 @ 1

5.3 Aeration valve

The RVC is equipped with an electromagnetic aeration valve. The rotor
chamber is aerated through this valve after the end of the evaporation
process.

NOTE

Unpressurised inert gas can also be used for aerating the rotor chamber.

Pos: 95 / 010 Unive rsalm odul e/ L ee rzeile @ 0\ mod _120 211 624 450 0_0. docx @ 11 4 @ @ 1

Pos: 96 / 200 Christ /37 1 RVC-BA (PROJEKTE) /RVC 2-2 5 CDpl us/05 0 Aufst ellun g un d Anschl uss/0 50- 003 0-0 010 Opti on: Bel üftu ngs- und Luf tinjektio nsve ntil (Ve ntilblock ) @ 8\m od_1 315 901 791 424 _68. docx @ 50 497 @ 4 @ 1

5.3.1

Option: Aeration and micro injection valve (Valve block)

As an option, the rotational vacuum concentrator can be equipped with an
electromagnetic aeration and injection valve (part no. 125936).
In addition to the functions of the standard valve, this valve can also inject
air. If the vacuum value falls below the preset value, this valve opens briefly
to ensure a constant vacuum.

NOTE

If the valve block is installed, the configuration of the control unit must be

adapted accordingly (see chapter 6.5.3.5 - "Options").




1

Fastener for non-required

connection element


2

Air injection


3

Aeration



Fig. 7: Valve block (option)

NOTE

Just like the standard aeration valve, the valve block can also use
unpressurised inert gas.