7. Interrupting an evaporation process:

• Press the left-hand function key "mode".
• Select "Stop evaporation" and confirm.
Without an interruption, the evaporation process ends after the
preselected time. The following process runs automatically:

• At maximum speed: the stop valve closes the connection to the vacuum

pump, the aeration valve opens; the rotor speed is maintained for 30
sec (aeration phase, to avoid delays in boiling).

• Rotor decelerates until it is at a standstill. After the standstill of the rotor

the lid can be opened in order to remove the rotors.


Function and control elements:

1 Lid
2 Rotor chamber
3 Control panel
4 Mains power switch
5 Lid lock device
6 Rotor shaft
7 Name plate
8 Electrical vacuum measuring

sensor connection

9 Option: serial interface RS 232
10 Power supply connection of

the vacuum pump

11 Power supply connection of

the pressure control valve

12 Power supply connection of

the stop valve

13 Equipotential bonding screw
14 Mains connection and mains

fuse protection

15 Vacuum connection
16 Connection for aeration
17 Option: Valve block

Fig. 59: Total view of 2-25 CD

plus

Abb. 60: Rear view of 2-25 CD

plus

Pos: 221 /01 0 Univ ersal mod ule/ Sei tenwe chsel @ 0\ mod _12 021 1624 431 2_0 .docx @ 1 05 @ @ 1