Martin Christ RVC 2-25 CDplus User Manual

Page 47

Advertising
background image

RVC 2-25 CD&

6 Operation

Version 11/2008, Rev. 2.2 of 04/12/2013 • sb

47

Translation of the original operating manual

6.5.3.3 Special functions

Pressure increase test

NOTE

The pressure increase test can only be performed if a vacuum sensor is
installed.


Entering the set values
• Select "Special functions" from the main menu.
• Open the menu "Pressure increase test" and select "Parameter

pressure increase test".

• Enter the set values, confirm and quit the menu.

Performing the pressure increase test
• Select "Special functions" from the main menu.
• Open the menu "Pressure increase test" and select "Start pressure

increase test".

The pressure increase test will be performed. The test time progress and
the measured pressure increase are displayed in a graphical form.

Fig. 39: Graphical presentation of the pressure increase test


When the test is finished, an information window pops up and shows the
result of the pressure increase test.

Pos: 149 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1

Pos: 150 /20 0 Ch rist/3 70 RVC-BA (S TANDARDMODULE) /06 0 Betri eb/CDpl us/06 0- 0050 -00 30 -004 0 Pr ozess - un d Anla gen -Inf orm ation en @ 8\ mod _130 993 318 443 3_6 8.doc x @ 4 711 5 @ 4 @ 1

6.5.3.4 Process and equipment information

The "Process & equipment Information " menu informs the user about any
error, process, or system messages.
In the event of a message, a sound signal can be heard and the symbol
" " is displayed on the status bar. In addition, the process and equipment
information is displayed. If the user is in a menu, the window will not be
displayed until the user quits the main menu.

1

Message

2

Status of the

Information

3

Number of messages

pending

4

Function key for

quitting the menu

5

Function key for
acknowledging the
message

Fig. 40: Structure of the process & equipment info window