Other features and settings – SOR Calibration Manager Software User Manual

Page 11

Advertising
background image

11/12

Form 1459 (01.13) ©SOR Inc.

Re-Zero Feature

The Manual Zero Adjust feature can be used to adjust the Zero

Set Point to compensate for Zero Set Point shifts associated

with environmental effects. For example, a pressure transmit-

ter that was calibrated at the Lenexa, KS (USA) factory may

read less than 4mA at zero pressure when placed in service at a

high-altitude location. To correct this, the output at zero can be

adjusted back to the factory calibrated zero output.

NOTE: In order for the Manual Zero Adjust feature to be enabled for use, the device

must be successfully calibrated prior to usage of this feature. If a calibration has

been started, but not successfully completed, the Manual Zero Adjust feature will

not be available until a successful calibration has been completed.

Other Features and Settings

Changing the Display units

The display units for the application can be changed to English Units or Metric Units

for convenience. To change the display units, go to View on the menu bar and select

the preferred display units.

Obtaining Software Version Information

To view application information regarding SOR Inc., application revision, etc, go to

Help on the menu bar and click “About .”

Warranty

The Calibration Manager software and

the Calibration Kit hardware are covered

by SOR warranty for six months. For

more information please contact

Customer Service.