VERIFyIng ThE CalIBRaTIOn

To verify the calibration status of a

device, begin by clicking the “Verify

calibration” button. Follow the guided

steps to complete the verification. If the

verification results are not acceptable,

start the “Calibration wizard” to either

adjust the calibration or re-range the

device.

NOTE: It may be desirable to save an

archive of the device calibration when the

device has been calibrated to a satisfactory

zero and span output. To do this, press the

“Save Report” button. You will be prompted

for a directory to save the file. Once the

directory is chosen, it will automatically be

recalled upon subsequent saves.

After the “Finish” button is pressed, the following

confirmation will be displayed to confirm the

calibration process was completed successfully.