SOR Calibration Manager Software User Manual

Page 6

Advertising
background image

6/12

Form 1459 (01.13) ©SOR Inc.

Calibration

To calibrate a device, begin by clicking

the “Calibration wizard” button. Simply

follow the guided steps to complete the

calibration. New Zero & Span Set Points

can be entered to re-range the device if

needed.

The guided calibration wizard will

provide step-by-step instructions on

how to calibrate the device. Follow these

instructions to guide you through the

calibration process.

NOTE: Take care to ensure the USB cable

remains connected during calibration. If the

USB cable should become disconnected at

any point during the calibration, it may be

necessary to exit the Calibration Manager
and restart the Calibration Wizard.

After the first attempt to calibrate a

device is completed, the summary

screen is displayed to show how

closely the device is calibrated to the

target settings.

Press the “Next” button to further adjust

the Zero and Span Set Points. Once the

device is calibrated to a satisfactory

precision, press the “Finish” button to

complete the calibration.

NOTE: It may be desirable to save an

archive of the device calibration when the

device has been calibrated to a satisfactory

zero and span output. To do this, press the

“Save Report” button. You will be prompted

for a directory to save the file. Once the

directory is chosen, it will automatically be

recalled upon subsequent saves.