14)
The recommended windscreen has negligible effect at frequen-
cies <12.5kHz.
15.
The instrument should be calibrated with either a CEL-284/2
electroacoustic calibrator or an RFT 05001 pistonphone.
16.
For maximum accuracy the operator and instrument case
should be more than 1m behind the microphone.
17.
The CEL-393 is designed to be used only with its own internal
filter set. Models containing filters are designated CEL-393B
and CEL-393D.
18.
The electrical impedance of equipment attached to the log DC
or conditioned AC output connectors should be >3k3 Ohms.
Digital outputs are available for CEL-238 Secondary Proces-
sor or CEL-433 Serial Interface.
19.
The reference frequency is 1kHz.
20.
The reference range is '70' ('60' on C and D versions).
/90.
393 Handbook