You can easily review any or all of the measurements that you have "Saved." To view the
currently opened Job File, Press the "File" menu and then "View Job File.."

You will now see the data of the last measurement taken. Press the up arrow on the scroll bar to
see the very top line of the last measurement.

It will be similar to the following:

=Ivie STI-PA v1.0===MV38.8===
Name of Job File
Area
Location 001
19 April 13:27 hours

Single Test Result
STI-PA-0.69 CIS-0.84 Leq-51.0

Band LEQ MF1 MF2 Status Level
8000 Hz 82.1 0.09 0.06 ok 40.1
4000 Hz 82.0 0.05 0.03 ok 68.3
2000 Hz 78.5 0.06 0.16 ok 73.7
1000 Hz 74.3 0.17 0.18 ok 76.8
500 Hz 76.9 0.11 0.09 ok 89.7
250 Hz 71.3 0.14 0.20 ok 90.5
125 Hz 73.8 0.28 0.20 ok 81.0
------------------------------

Explanation by Line:

Line #1
=Ivie STI-PA v1.0===MV38.8===
v1.0- is the version number
MV- Male voice was selected
38.8- is the Calibration constant

Line #2
Name of Job File-Name of Job file.

Line #3
Area - Area description

Line #4
Location - Location number in the area