88

Table 2
For all ME EQUIPMENT and ME SYSTEMS

Guidance and manufacturer’s declaration – electromagnetic immunity

The BC 31 is intended for use in the electromagnetic environment specifi ed below.
The customer or the user of the BC 31 should assure that it is used in such an environment.

IMMUNITY test

IEC 60601 test level Compliance level Electromagnetic environment – guidance

Electrostatic discharge (ESD)
IEC 61000-4-2

± 6 kV contact
± 8 kV air

± 6 kV contact
± 8 kV air

Floors should be wood, concrete or ceramic tile. If
fl oors are covered with synthetic material, the relative
humidity should be at least 30 %.

Power frequency (50/60 Hz)
magnetic fi eld IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fi elds should be at levels
characteristic of a typical location in a typical commer-
cial or hospital environment.

NOTE: U

T

is the a.c. mains voltage prior to application of the test level.