• the batteries are almost empty

.

In such cases, repeat the measurement. Ensure you do not

move or speak.

If necessary, re-insert or replace the batteries.
Technical alarm – description
Should the recorded blood pressure (systolic or diastolic)

lie outside the limits specified in the section “Technical

specifications”, the technical alarm will appear on the dis-

play indicating either “

Hi” or “Lo”. In such cases, you should

seek medical assistance and check the accuracy of your

procedure.

The limit values for the technical alarm are factory set and

cannot be adjusted or deactivated. These alarm limit val-

ues are accorded second priority under the standard IEC

60601-1-8.

The technical alarm is a non-locking alarm and must not be

reset. The signal shown on the display will disappear auto-

matically after about 8 seconds.

7. Cleaning and care

• Clean your blood pressure monitor carefully using a

slightly damp cloth only.

• Do not use any detergents or solvents.

• Never hold the instrument under water as otherwise liquid

can penetrate and damage the instrument.

• Never place any heavy objects on the instrument.

8. Technical details

Model no.

BC 44

Measurement

method

Oscillometric, non-invasive blood pres-

sure measurement on the wrist

Measurement

range

Cuff pressure 0 – 300 mmHg,

systolic 60 – 260 mmHg,

diastolic 40 –199 mmHg,

Pulse 40 –180 beats/minute

Display accuracy Systolic ± 3 mmHg, diastolic ± 3 mmHg,

pulse ± 5 % of the value shown

Measurement

inaccuracy

Max. permissible standard deviation

according to clinical testing:

systolic 8 mmHg /diastolic 8 mmHg

Dimensions

L 87,8 mm x W 65,8 mm x H 28,6 mm

Weight

Approx. 98 g (without batteries)

Cuff size

140 to 195 mm

Permissible oper-

ating conditions

+ 10 °C to + 40 °C, ≤ 90 % relative air

humidity (non-condensing)

Permissible stor-

age conditions

- 20 °C to + 55 °C, ≤ 90 % relative air hu-

midity, 800 –1050 hPa ambient pressure

Power supply

2 x 1,5 V

AAA batteries