Electromagnetic compatibility information – Beurer BC 58 User Manual

Page 106

Advertising
background image

106

Electromagnetic Compatibility Information

Table 1

For all ME EQUIPMENT and ME SYSTEMS

Guidance and manufacture’s declaration – electromagnetic emissions
The BC 58 is intended for use in the electromagnetic environment specified below.

The customer or the user of the BC 58 should assure that it is used in such an environment.
Emissions test

Compliance

Electromagnetic environment – guidance

RF emissions CISPR 11

Group 1

The BC 58 uses RF energy only for its internal function. Therefore, its RF emissi-

ons are very low and are not likely to cause any interference in nearby electronic

equipment.

RF emissions CISPR 11

Class B

The BC 58 is suitable for use in all establishments other than domestic and

those directly connected to the public low-voltage power supply network that

supplies buildings used for domestic purposes.

Harmonic emissions IEC 61000-3-2

Not applicable

Voltage fluctuations/flicker emissions

IEC 61000-3-3

Not applicable

Table 2

For all ME EQUIPMENT and ME SYSTEMS

Guidance and manufacturer’s declaration – electromagnetic immunity
The BC 58 is intended for use in the electromagnetic environment specified below. The customer or the user of the BC 58 should assure

that it is used in such an environment.
IMMUNITY test

IEC 60601

test level

Compliance

level

Electromagnetic environment – guidance

Electrostatic discharge (ESD)

IEC 61000-4-2

± 6 kV contact

± 8 kV air

± 6 kV contact

± 8 kV air

Floors should be wood, concrete or ceramic tile. If floors are covered

with synthetic material, the relative humidity should be at least 30 %.

Power frequency (50/60 Hz)

magnetic field IEC 61000-4-8 3 A/m

3 A/m

Power frequency magnetic fields should be at levels characteristic of

a typical location in a typical commercial or hospital environment.

NOTE: U

T

is the a.c. mains voltage prior to application of the test level.